Emergency Power
- Integrated 2hr battery backup
Alarm functions
- Event/alarm log on display
- Humidification system alarm
- Alarms can be muted during servicing
- Bottle replacement reminder alarm
Remote monitoring
- Connectivity to clinic network
Password protection
- Prevents accidental power off
- Prevents unauthorised parameter change
Humidification system
- Works with premixed gas
- Minimal tubing to mitigate kinking and gas reduction
- Rapid recovery to optimal temperature, gas and humidity
- Improved user-friendly bottle design
- Easier bottle replacement
Service friendly
- Easy hatch opening for battery replacement
- Easy filter replacement/cover removal
- Service is sold separately and recommended
Full contact plates
- Consistent heat distribution
- Absolute control to 0.1°C
- Seven point temperature control
- Large range of culture dish sizes accepted
Product Specifications
The BT37 Mark II has been developed in conjunction with PLANER™; all trademarks acknowledged. Specifications may change.
| System | |
| Dimensions (WxDxH) | 435 mm wide x 330 mm deep x 185 mm high |
| Weight | 17 kg |
| Storage temperature | -10°C to +50°C |
| Storage humidity | 5% to 95% relative humidity non-condensing |
| Storage special instructions | Recharge every 4 months by connecting to the mains power supply for 24 hours |
| Capacity | |
| Dishes per chamber | 4 x Nunc® 4 well Petri dishes 10 x Nunc 30 mm Petri dishes 4 x Minitub 5 well Petri dishes 4 x Falcon® 60 mm Petri dishes |
| Internal Battery | |
| Internal battery backup | 2 hour backup |
| Control | |
| Temperature control range | (ambient + 5°C) to (ambient + 20°C) 40°C max.> |
| Temperature measurement accuracy | ± 0.2°C |
| Temperature control accuracy | ± 0.1°C measured after any transient effects due to set-point changes have subsided |
| Flow control range | 0 ml/minute to 900 mL/minute. Normalized to 0°C, 50% RH and 1 bar |
| Flow Accuracy | The greater of ± 10% or ± 3 ml/minute |
| Flow control accuracy | The greater of ± 5% or ± 2 ml/minute measured after any transient effects due to set-point changes have subsided |
| Gas Supply | |
| Gas supply | Premixed gas. Typically 6% CO2, 5% O2, 89% N2 |
| Supply pressure | 1.5 ± 0.15 bar |
| Connectors | SWAGELOK 1/4” tube fitting |
| Monitoring | |
| Local area network (LAN) | 10 Base T Ethernet – RJ45 shielded Modbus-TCP-IP protocol |
| Independent temperature monitoring | Independent sensors can be fitted to the monitoring ports Recommended sensor type: PT100 Class A to EN60751 Maximum diameter: 2.51 mm |
| Power | |
| Power requirements | 100 – 240 V~, 50/60Hz, 2 A |







Reviews